Product ID: SK200, SK200C
HighLife HME (FOR USE WITH VENITILATOR/RESPIRATOR)
Single patient use on anesthetized patients and respiratory care patients who require a breathing circuit. The product is intended to be replaced at least once every 24 hours.
|


|
Specifications:
- SK200
- Measure Output temperature @2hrs, 10breaths/min (ISO 9360)
- Vt250ml: >30°C, Vt500ml: >30°C, Vt750ml: >30°C, Vt1000ml: >30°C
- Resistance to flow(ISO 9360): 15/30/60/min, <0.5/1.1/2.1cm H2O
- Dead space (connectors included-ISO 9360): 34ml
- Weight: 22g
- Absolute Humidity (500ml Vt, 20BPM): >29mg H2O/L
- Tidal Volume: 250-1500ml
- Sterile: Gamma/E.O GAS
- Patient: Adult
- SK200C
- Measure Output temperature @2hrs, 10breaths/min (ISO 9360)
- Vt250ml: >32°C, Vt500ml: >32°C, Vt750ml: >32°C, Vt1000ml: >32°C
- Resistance to flow(ISO 9360): 15/30/60/min, <1.0/2.5/3.5cm H2O
- Dead space (connectors included-ISO 9360): 32ml
Features:
- SK200
- Hygroscopic HME media traps patient's own moisture, and minimizes heat loss.
- Low air flow resistance.
- Moisture output greater than 30mg H2O/I (ISO 9360, Vt750ml).
- Calcium chloride treated HME media improves effectiveness.
- ISO standard 15 and 22mm fittings.
- Patient end.
- SK200C
- Hygroscopic HME media traps patient's own moisture, and minimizes heat loss.
- Low air flow resistance.
- Moisture output greater than 30mg H2O/I (ISO 9360, Vt750ml).
- Calcium chloride treated HME media improves effectiveness.
- ISO standard 15 and 22mm fittings.
- Patient / Machine end.
- Bacterial and Virus Filtration efficiency of filter media are tested and Performed at Nelson Laboratories, Inc. according to procedure No. 48061 for BFE and Procedure No. 920694-1, Laboratory No. 48060 for VFE.
Certificate:
|
Ordering information:
- Minimum Order: Negotiable
- TEL: 886-2-86782106-7
- FAX: 886-2-86782108
|
|